Differential Pressure Smart Manometer psid
Features NIST traceable accuracy of ±0.05% f.s. independent of temperature effect Leak test function displays min/max values until user-selected test time expires and then displays the leak rate per minute Min/Max function captures the extremes of pulsating pressure signals Exponential damping with a user-selectable time constant from 0.1 to 25 seconds minimizes pulsation effects Field calibration of zero, span and linearity (user has the option to select a 1-point or 5-point calibration procedure) Hold function freezes the display Auto Record documents up to 240 reading for future recall All pressure ranges are rated for 1000 psi common mode pressure HI side overpressure rating is 3X f.s., while the LO side overpressure rating is 3X f.s. or 150 psi, whichever is less Optional flushing ports allow solvent cleanout of P1 and P2 ports or can be used to vent air from the connecting tubing of liquid applications 1/8" female NPT 316L SS HI and LO ports, 5/16-24 SAE/MS J1926 flushing ports with 316L SS plugs Wetted parts are 316L SS with Viton O-rings 5-digit LCD (0.25" high), 2-line x 16-alphanumeric characters with contrast adjustment Two-stage backlight Operates continuously for 100 hours on 4 "AA" batteries (included) User-selectable shut-off is programmable at 10, 20, 30, 45, or 60 minutes Polycarbonate, permanently dissipative base unit with sealed membrane keypad
EGP190,002